Physical Property Measurement System

Equipment & Details

DynaCool – 14T Cryogen Free PPMS Base system
P.O. No.: 02/IIT/PO/F/EQ-02/CRF/2025 Dt. 17-07-2025
Make : Quantum Design
Model : PPMS DynaCool – 14T

Sample Details

Important Note
1.    Sample contacts should be prepared by the user.
2.    Please come one day before the scheduled measurement slot, as the sample needs
3.    to be mounted in the puck and allowed to dry properly before measurement.
4.    The voltage measurement range is 44 μV–4 V (resistivity 50µΩ - 1MΩ) for four probe measurements and 10 mV–10 V (2MΩ - 1GΩ) for two-probe measurements.


Rate : 














Utility and Working Principle

Temperature

1.    Measurable temperature range: 4–380 K
2.    Maximum temperature sweep rate: 10 K/min
3.    Preferred sweep rate during data acquisition: 4–5 K/min
4.    Below 10 K, a slower sweep rate of approximately 1–2 K/min is preferred.

Magnetic Field
1.    Measurable magnetic-field range: +14 T to −14 T
2.    Maximum field sweep rate: 100 Oe/s


Approximate Measurement Time
1.    R(T) measurement (4–380 K): approximately 6 hours for each scan
2.    R(H) measurement: approximately 3 hours at each temperature (1 quadrant)
3.    Sample length less than 10mm (preferable 5-10mm)